The Biomedical Sciences Flow Cytometry Core Laboratory uses a BD-Biosciences FACS Aria high speed flow
cytometer/cell sorter utilizing a quartz cuvette for interrogation. It is equipped with three diode lasers providing excitation
lines at 488 nm, 633 nm and 407 nm. Signal processing is digital,
allowing high acquisition rates, recapitulation and logical analysis of acquired data. There are 5 emission wavelengths detected
from 488 nm excitation, 3 from 633 nm excitation and 3 from 407 nm excitation to accommodate a wide range
of fluorochromes. The Aria is capable of high speed cell sorting (>30,000 events/second) and collecting sorted samples
into up to four 2ml or 5ml tubes, two 15ml tubes
and is also equipped with a robotic device for sorting into multi-well plates and onto glass slides.
| FACSAria Cell Sorter (Current Configuration) | |||||
| Design of the BD octagon/trigon collection Optics offer greater flexibility than was available in previous instruments. If you need different filter combinations, please ask about different configurations. | |||||
| Lasers | Filters | ||||
| 488nm Argon | FITC 515-545nm |
PE 564-606nm |
PE TEXAS RED 600-620nm |
PE/PerCP/Cy-5 675-715nm |
PE/Cy-7 750-810nm |
| 633nm HeNe | APC 650-670nm |
APC/Cy-7 750-810nm |
  |   |   |
| 407nm Violet | Violet1 ALEXA-430 515-545nm |
Violet2 DAPI/HOECHST/PacificBlue 430-470nm |
  |   |   |
The Biomedical Sciences Flow Cytometry Core Laboratory has access to a BD-Biosciences LSRII Flow Cytometer also
utilizing a quartz cuvette for interrogation. It is equipped with three lasers providing excitation
lines at 488nm, 633nm and the UV line at 325nm. Signal processing is digital,
allowing high acquisition rates, recapitulation and logical analysis of acquired data. There are 4 emission wavelengths detected
from 488 nm excitation, 2 from 633 nm excitation and 2 from 407 nm excitation to accommodate a wide range
of fluorochromes.
| LSRII (Current Configuration) | |||||
| Design of the BD octagon/trigon collection Optics offer greater flexibility than was available in previous instruments. If you need different filter combinations, please ask about different configurations. | |||||
| Lasers | Filters | ||||
| 488nm Argon | FITC 530/30nm |
PI 610/20nm |
PE 576/26nm |
PE/PerCP/Cy-5 695/40nm |
|
| 633nm HeNe | APC 650-670nm |
APC/Cy-7 750-810nm |
  |   | |
| UV-Currently unavailable | Offline ------- 515-545nm |
Offline ------- 430-470nm |
  |   | |